Tenders Description & Specification for Inductively Coupled Plasma Spectrometer
Inductively Coupled Plasma Spectrometer
Inductively Coupled Plasma Spectrometers can detect approximately 70 trace to constant elements in different materials. They are widely used in scientific research Institutes, colleges, and other industry enterprises related to the fields such as geology, metallurgy, rare earth materials, electro-plating, cement, petroleum, chemical engineering, environmental detection, non-ferrous metals, medical science, food and agriculture.
Performance characteristics
Simultaneous measurement of multi elements
Low detection limit: 0.1-10ppb, and even lower
Wide density linear range, 5-6 orders of magnitude
Less chemical interference, maximum accuracy, RSD≤2%
High analysis speed,low cost
Technical specifications
RF Generator Specifications
Circuit type: inductively feedback self-oscillation circuit, coaxial cable transmission, matched tuning, power feedback close-loop auto-control
Working frequency: 40MHz±0.05%
Frequency stability: <0.1%
Power output: 800W-1200W
Power output stability: <0.3%
Intensity of electromagnetic leakage: electric filed intensity in 30cm distance to the body: E:<2V/m
High Frequency Box
High Frequency Box excites very stable plasma torch by exploiting the inductively feedback self-oscillation circuit. The circuit is transmitted by coaxial cable, matchingly tuned and implements close-loop auto-control by means of power feedback (National Patented Invention). Industrial argon can be used in this kind of generator, which not only makes the lighting the ICP torch easy but also saves the cost of highly purified argon.
Match box
The high performance match box is composed of vacuum ceramic adjustable capacitors and gas adjustable capacitors, which allows it to reach the best match state and achieve maximum effective output on load coils.
Directional coupler
The stability of power output is achieved by controlling the anode voltage of oscillation valve after the power signals are sent out of the directional coupler and get amplified by the amplifier.
Sample introduction system specifications:
Inner diameter of output coils: 25 mm, 3 coils
Torch: three concentric torches, outer diameter 20 mm, quartz
Outer diameter of coaxial nebulizer: 6mm
Outer diameter of double-pass spray chamber: 34mm
Specifications of Ar flowmeter and carrier gas pressure gauge:
Plasma gas flowmeter: (100-1000)L/h (1.6-16L/min)
Auxiliary gas flowmeter(10-100)L/h (0.16-1.66L/min)
Carrier gas flowmeter10-100)L/h (0.16-1.66L/min)
Carrier gas pressure stabilization valve (0-0.4MPa)
Cooling water: Water temperature: 15-25℃ Flow rate >5L/min Water pressure: >0.1MPa Resistivity>1MΩ
Technical Specifications of Optical System
Optical path: Czerny-Turner
Focal length: 1000mm
Grating specification: Ion Beam Etching Holographic Grating; The etching density is 3600 grooves/mm; the etching area is 80*110 mm.
(Alternative 2400 grooves/mm; etching area: 80*110mm)
Reciprocal linear dispersion rate: 0.26nm
Resolution: ≤ 0.008nm
Scan wavelength range:
3600 grooves/mm scan wavelength range: 195—500 nm
2400 grooves/mm scan wavelength range: 195—800 nm
Minimum pace of stepper motor drive: 0.0006nm
Exit and entrance slit: 20μm
Speculum specifications: (78×105×16)mm
Lensφ30,1:1 imaging
Thermostatic device for optical system: 32℃±1℃
Ltekonline are the manufacturers of Inductively Coupled Plasma Spectrometer used to demonstrate engineering principles in Engineering Colleges, Engineering Schools and Petroleum Testing Lab for Engineering Teaching Lab in Universities
Categories: Petroleum Testing Lab