Tenders Description & Specification for Neonatal Infant Incubator for Baby
Neonatal Infant Incubator for Baby
Standard Configuration:
Neonate bilirubin phototherapy equipment
The hood, the base, the controller, I.V pole, shelf.
Specification:
Power Output: 850VA
Temperature Variability: <= 0.5°C
Temperature Uniformity: <= 0.8°C
Mode of control: Air mode control by micro-computer
Air temperature control range: 25°C - 37°C
Power supply: AC22—230V/50Hz or AC110V-120V/50-60Hz
Noise inside hood: <=55Db (A) (Environment noise<=45dB (A))
Optional with DC motor. Noise inside hood: <=45Db (A) (Environment noise<=35dB (A))
Failure alarm: over temperature alarm, deviation alarm, sensor failure alarm, fan motor alarm, power failure alarm, system alarm and so on.
Optional Configuration:
The light source of the fixed neonate bilirubin
Phototherapy equipment is LED bulb
Skin temperature control and disposable skin temperature sensor
DC motor
>37°C override mode
Environmental Temperature:
Operating range: 20°C ~ 30°C
Ambient air movement rate: <0.3m/s.
Ltekonline are the manufacturers of Neonatal Infant Incubator for Baby used to demonstrate engineering principles in Engineering Colleges, Engineering Schools and Infant Incubator for Engineering Teaching Lab in Universities